AO-06 Medical Oxygen Sensor O2 Cell MOX-4 Replacement Oxygen Battery for Aeonmed Ventilator Anesthesia Machine
Support Payment Term:
PayPal
USD
EUR
GBP
SGD
CNY
MXN
BRL
JPY
THB
AUD
$37.50
Discover similar items
Page 1 of 2
Technical Specification:
Measurement | Operating Principle | Partial pressure electrochemical |
Output Voltage | 9 ~ 13 mV (in ambient air, 20.9% O₂) | |
Measurement Range | 0 ~ 100% O₂ | |
Response Time (T90) | < 15 seconds | |
Linearity | Linear 0% to 100% O₂ | |
Electrical | Temperature Compensation | < 2% O₂ equivalent (0 ~ 40 °C) |
External Load Resistor | ≥ 10 kΩ | |
Mechanical | Housing Material | White medical-grade ABS |
Weight | 40 g (nominal) | |
Environmental | Operating Temperature | 0 ~ 50 °C |
Operating Pressure | 0.5 ~ 2.0 bar | |
Operating Humidity | 0 ~ 99% RH (non-condensing) | |
Performance | Long-term Output Drift | < 5% voltage loss per year (in 100% O₂) |
Expected Service Life | 1.5 × 10⁶ %O₂·hours @ 20 °C; 0.8 × 10⁶ %O₂·hours @ 40 °C |