About this item:

1. Flat panel detector: Material: amorphous silicon cesium iodide implant size: 17 × 17 inch pixel: 9 million pixel A / D: 16bits DQE: 65% pixel matrix: 3072 × 2072 pixel size: 140 μ m imaging time: < 3S limit resolution: ≥ 3.5lp/mm data transmission mode:Mobile wired 2. tube: focus size: 0.6 / 1.2 focus power: 20kW / 50kw-50hz anode heat capacity: 150khu heat capacity of tube sleeve: 1250khu anode target angle: 12 ° maximum tube voltage: 150kV 3. High Voltage Generator Power :50 KW

Support Payment Term:

PayPal

USD

EUR

GBP

SGD

CNY

MXN

BRL

JPY

THB

AUD

More

$82999.00

Quantity
Price not good?Product not right?Need more support?Chat here!

Discover similar items

Page 1 of 2

1. Flat panel detector:

Material: amorphous silicon cesium iodide implant

size: 17 × 17 inch

pixel: 9 million pixel

A / D: 16bits

DQE: 65% pixel

matrix: 3072 × 2072 

pixel size: 140 μ m 

imaging time: < 3S

limit resolution: ≥ 3.5lp/mm

data transmission mode:Mobile wired

 2. tube:

focus size: 0.6 / 1.2

focus power: 20kW / 50kw-50hz 

anode heat capacity: 150khu 

heat capacity of tube sleeve: 1250khu 

anode target angle: 12 ° 

maximum tube voltage: 150kV

3. High Voltage Generator Power :50 KW

Photographic KV: 40kV~150 kV

Photography mA: 10~ mA 630

mAs: 1-630

inverter frequency :> 40kHz

ARP:2400 Data

AEC: support

Control mode: workstation synchronous control

4. Beam limiter

Specification: AC24V / 100W

Control mode: Manual

Minimum irradiation field: < 50 mm × 50 mm

Radiation Vision(sid = 1000mm): 430mm × 430mm

Average illumination brightness:> 140lux;

Contrast: > 4:1

Inherent filtering: 1.0 MMAL equivalent; Lead equivalent of shield blade: 3 mmpb (middle blade 2 mmpb)

5. Inlet grid

Grid density: ≥ 80L / cm

Grid ratio: ≥ 10:1

Grid focal length: 100cm / 180cm

6. Dual column multi-functional photography system

● double column supporting structure

●Manual vertical / horizontal position conversion (electric auxiliary)

●Projection position: standing position, horizontal lying position and oblique position

●X tube lateral movement range: lateral movement stroke 1800mm ± 50mm

●Vertical movement range of X-tube: vertical travel 1500mm ± 50mm

●Rotation range of X-tube: rotation angle 0 °± 2 ° ~ ± 90 °± 2 °

●Vertical movement range of detector bracket of chest X-ray stand: 500mm ~ 1700mm ± 5%

●Standing film distance SID: 1000mm ~ 1800mm

●Lying film distance SID: 1000mm

●Bed structure

●Four phase floating bed

●Bed surface: length × width 2200mm × 800mm

●Material: Polymer carbon fiber

7. Diagnosis workstation

CPU: cool capacity i5 quad core 3.2ghz

RAM: 4GB

Hard disk capacity: 1TB

Optical drive: DVD drive

System interface: USB, RS232, DVI / VGA / HDMI / DP, com serial port, PCI / pci-e

Display: 22 "LCD

Operating system: genuine Windows 7 Professional Edition

Diagnostic Workstation 

Workstation software function

User operation: patient information registration, Multi condition query of some patients' information, patient appointment, emergency

Image processing functions: window width / window level, local automatic window width / window level, preset window width / window level, positive and negative image inversion, image flip, rotation, image enlargement and roaming, image interpolation edge enhancement, local magnification, restoration of original image, text / number annotation, image ruler line segment measurement, angle measurement, brightness / contrast adjustment

Image printing: DICOM printing, paper printing, image contrast / brightness adjustment, multiple image splicing printing, film size selection, print queue control