direct digital X-ray photography system DX-3000B
About this item:
Support Payment Term:
PayPal
USD
EUR
GBP
SGD
CNY
MXN
BRL
JPY
THB
AUD
$68999.00
Discover similar items
Page 1 of 2
1. Flat panel detector:
Imaging area: ≥ 17 × 17 inches (43cm × 43cm)
Detector material: amorphous silicon cesium iodide
Pixel sum: ≥ 9.4 million
Acquisition matrix: 3K × 3K(3072 × 3072)
Data transmission mode: wireless transmission
Pixel size: ≤ 139 μ M
Resolution: ≥ 3.6lp/mm
DQE:≥70%
Imaging time: ≤ 10s
A / D conversion: ≥ 16bit
Storage temperature: - 15 ℃~ 55 ℃, service temperature: 10 ℃~ 40 ℃
Flat panel detector with noise elimination technology
Calibration mode: offset calibration, gain calibration, bad pixel calibration, line noise calibration
2. Ball tube:
Tube focus: ≤ 0.6 mm / 1.2 mm
Focus power: ≥ 100kW
Anode heat capacity: ≥ 380khu
Anode target angle: 12 ° Voltage range: 40kv-150kv
Anode speed: ≥ 9500rpm
Protection device: oil pressure protection switch
3. High voltage generator
Maximum output power: 75KW
Inverter frequency: 200kHz
Output voltage: 40kV - 150kV
Maximum output Ma: ≥ 900mA
Maximum current time product: ≥ 900mas
Minimum exposure time: ≤ 1ms
Increment: ≤ 1kV regulation
Exposure mode:
Manual setting of parameters exposure mode, organ program photography mode
The photography method should be as follows:
General photography, filter photography
Power demand: 380V three phase power input
With exposure protection control device
With fault status display function
Network integration: the generator parameter status can be integrated with the collected image display
With high voltage generator shutdown protection device
4. Beam limiter
Specification: AC24 V / 100 W
Operation mode: Manual
Minimum field: < 50mm × 50mm
Radiation field (sid = 1000mm): 430mm × 430mm
Average illumination intensity: > 140lux;
Illumination ratio: > 4:1
Inherent filtration: 1.0 mmAL equivalent; Lead equivalent of shield blade: 3 MMPB (2 MMPB for middle blade)
Brightness indication time of beamer: ≥ 30 seconds
5. Inlet screen
Grid density: ≥ 80L / cm
Grid ratio: ≥ 10:1
Grating focal length: 100cm / 180cm
6. Double column multi-functional photography system
●Double column support structure
●Manual vertical / horizontal position conversion (electric auxiliary)
●Projection position: standing position, horizontal lying position and oblique position
●Transverse movement range of X-tube: 1800mm ± 50mm
●Vertical movement range of X-tube: vertical movement stroke 1500mm ± 50mm
●Rotation range of X-tube: rotation angle 0 °± 2 °~±~90°± 2 ° ●Vertical movement range of detector bracket of chest X-ray frame: 500mm ~ 1700mm ± 5%
●Stand film distance Sid: 1000mm ~ 1800mm
●The distance between film and Sid in supine position: 1000mm
● bed structure
●Four phase floating bed
●Bed surface: long × Width 2200mm × 800mm
7. Diagnostic workstation
CPU: coolong i5 4-core 3.2ghz
RAM: 4GB
Hard disk capacity: 1TB
Optical drive: DVD drive
System interface: USB, RS232, DVI / VGA / HDMI / DP, com serial port, PCI / pci-e
Display: 22 "LCD
Operating system: Windows 7 Professional Edition
Diagnostic Workstation
Workstation software function
User operation: patient information registration, multi condition query of location patient information, patient appointment, emergency department
Image processing functions: window width / window level, local automatic window width / window level, preset window width / window level, positive and negative image reversal, image reversal, rotation, image magnification and roaming, image interpolation edge enhancement, local magnification, restoration of original image, text / digital annotation, image ruler line measurement, angle measurement, brightness / contrast adjustment
Image printing: DICOM printing, paper printing, image contrast / brightness adjustment, multiple image combination printing, film size selection, print queue control
8. Medical image printer
Output film specification: 14 × 17、11 × 14、10 × 12、8 × 10 and other specifications
Output speed ≥ 80 sheets / hour
Resolution ≥ 320dpi
Film input ≥ 2 cassettes,
Gray scale resolution ≥ 14bit
DICOM3.0 interface